MODEL HMP35C
TEMPERATURE AND RELATIVE HUMIDITY PROBE
1. GENERAL DESCRIPTION
The HMP35C Temperature and Relative
Humidity probe contains a thermistor for
measuring temperature and a Vaisala
capacitive polymer H chip for measuring relative
humidity.
The -L option on the model HMP35C
Temperature and Relative Humidity probe
(HMP35C-L) indicates that the cable length is
user specified. This manual refers to the
sensor as the HMP35C.
2. SPECIFICATIONS
Operating Temperature: -35° to +60°C
Storage Temperature: -40° to +80°C
Probe Length: 25.4 cm (10 in.)
Probe Body Diameter: 2.5 cm (1 in.)
Filter: 0.2 µm Teflon membrane
Filter Diameter: 1.9 cm (0.75 in.)
Power Consumption: < 4 mA
Supply Voltage (via CSI switching circuit):
7 to 35 VDC
Settling Time after power is switched on:
0.15 seconds
2.1 TEMPERATURE SENSOR
Sensor: thermistor (100 k @ 25°C)
Temperature Measurement Range:
-35° to +55°C
Thermistor Interchangeability Error:
Typically < ±0.2°C over 0° to +60°C, ±0.4°C
@ -35°C
Polynomial Linearization Error:
< ±0.5°C over -35° to +50°C
2.1.1 Temperature Sensor Accuracy
The overall probe accuracy is a combination of
the thermistor's interchangeability specification,
the precision of the bridge resistors, and the
polynomial error. In a "worst case" all errors
add to an accuracy of ±0.4°C over the range of
-24°C to 48°C and ±0.9°C over the range of
-38° to 53°C. The major error component is the
interchangeability specification of the
thermistor, tabulated in Table 1. For the range
of 0° to 50°C the interchangeability error is
predominantly offset and can be determined
with a single point calibration. Compensation
can then be done with an offset entered in the
measurement instruction. The bridge resistors
are 0.1% tolerance with a 10 ppm temperature
coefficient. Polynomial errors are tabulated in
Table 2 and plotted in Figure 1.
TABLE 1. Thermistor Interchangeability
Temperature
(°C)
Temperature
Tolerance (±°C)
40
0.40
30
0.40
20
0.32
10
0.25
0 to +50 0.20
TABLE 2. Polynomial Error
Temperature
Range(°C)
Error (°C)
40 to +56 <±1.0°C
38 to +53 <±0.5°C
24 to +48 <±0.1°C
2.2 RELATIVE HUMIDITY SENSOR
Sensor: Vaisala capacitive polymer H chip
Relative Humidity Measurement Range:
0 to 100% non-condensing
RH Output Signal Range:
0.002 to 1 VDC
Accuracy at 20°C
±2% RH (0 to 90% Relative Humidity)
±3% RH (90 to 100% Relative Humidity)
Temperature Dependence of Relative Humidity
Measurement: ±0.04% RH/°C
Typical Long Term Stability:
Better than 1% RH per year
Response Time (at 20°C, 90% response to
steep change in humidity):
15 seconds with membrane filter