HMP35C TEMPERATURE AND RH PROBE
10
TABLE 9. Wiring for Example 4
Description Color CR10(X)
Temperature Orange SE 3 (2H)
Relative Humidity Green SE 4 (2L)
Temperature Signal Reference White AG
Temperature Excitation Black E1
Power Control Yellow C1
Power Red 12 V
RH Signal, Power, & Control Reference Purple AG
Shield Clear G
Example 4. Sample CR10(X) Program that Computes Vapor Pressure
and Saturation Vapor Pressure
;Measure the HMP35C temperature.
;
01: Temp (107) (P11)
1: 1 Reps
2: 3 SE Channel
;Orange wire (SE 3), White wire (AG)
3: 1 Excite all reps w/E1
;Black wire (EX 1)
4: 1 Loc [ T_C ]
5: 1 Mult
;See Table 4 for alternate multipliers
6: 0 Offset
;See Table 4 for alternate offsets
;Turn the HMP35C on.
;
02: Do (P86)
1: 41 Set Port 1 High
;Yellow wire (C1)
;Pause 150 milliseconds, before making the measurement, so the
;probe can stabilize on the true relative humidity.
;
03: Excitation with Delay (P22)
1: 1 Ex Channel
2: 0 Delay W/Ex (units = 0.01 sec)
3: 15 Delay After Ex (units = 0.01 sec)
4: 0 mV Excitation
;Measure the HMP35C relative humidity.
;
04: Volt (SE) (P1)
1: 1 Reps
2: 5 2500 mV Slow Range
;CR500 (2500 mV); CR23X (1000 mV); 21X, CR7 (5000 mV)
3: 4 SE Channel
;Green wire (SE 4), Purple wire (AG)
4: 2 Loc [ RH_frac ]
5: 0.001 Mult
;See Table 5 for alternate multipliers
6: 0 Offset
;Turn the HMP35C off.
;
05: Do (P86)
1: 51 Set Port 1 Low
;Yellow wire (C1)