Campbell-scientific Sentek EnviroSCAN EasyAG SDI-12 Manual v 3.4 Instrukcja Użytkownika Strona 20

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 30
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 19
EnviroSMART & EasyAG SDI-12 Probe Manual
Copyright © 1991 2003 Sentek Pty Ltd All rights reserved Page 15
Supported Commands
SDI-12 Commands Supported
The following commands are supported by the EnviroSMART and EasyAG SDI-12 probe interfaces:
Command Name Response
a!
Acknowledge
Active
a<CR><LF>
The probe interface address
aI!
Send
Identification
allccccccccmmmmmmvvvxxxxxxxxxxxx<CR><LF>
Identification information
aAb!
Change Address
a<CR><LF>
The probe interface address
aM!
Start
Measurement
atttn<CR><LF>
Delay (ttt) in seconds and number of
values (n) up to 9
aMn!
Start
Measurement
n in range “1”
- “3”
atttn<CR><LF>
Delay (ttt) in seconds and number of
values (n) up to 9
Note: A response of a0000<CR><LF> is
provided indicating all values are
served by previous Start Measurement
commands (aMn!)
aC!
Concurrent
Measurement
atttnn<CR><LF>
Delay (ttt) and number of values (nn) up
to 16
aCn!
Concurrent
Measurement
n in range “1”
- “9”
atttnn<CR><LF>
Delay (ttt) and number of values (nn) up
to 16
Note: A response of a0000<CR><LF> is
provided indicating all values are
served by previous Start Concurrent
Measurement (aCn!) commands
aDn!
Send Data
n in range “0”
- “9”
Refer to “Data Reading” section
aV!
Start
verification
a0000<CR><LF>
aRn!
Continuous
Measurement
a<CR><LF>
Not supported
Probe interface will respond with its
address
Przeglądanie stron 19
1 2 ... 15 16 17 18 19 20 21 22 23 24 25 ... 29 30

Komentarze do niniejszej Instrukcji

Brak uwag