CS420-L and CS425-L Pressure Transducers
Check indicating desiccant or enclosure humidity indicator; service
desiccant if necessary
Check battery condition (inspect physical appearance and use a keyboard
display or laptop to view the battery voltage)
Check all sensor readings (using keyboard display or laptop); adjust
transducer offsets if necessary
Check recent data using keyboard display or laptop
Perform routine maintenance suggested by manufacturers
See datalogger manual for more information about using a
keyboard display or laptop to view battery voltage, sensor
readings, and recent data.
NOTE
9.2 Every Three Months
Change batteries (as needed--may be less often)
Replace enclosure desiccants
Check calibration of all sensors
Inspect probe cable conditions for deterioration or damage
Check wire connections ensuring they are still secure
9.3 Every Two to Three Years or on a Rotating Schedule
Send the transducers to the factory or laboratory for inspection and have them
serviced and/or replaced as needed.
10. Troubleshooting
The most common causes of erroneous pressure transducer data include:
poor sensor connections to the datalogger
damaged cables
damaged transducers
moisture in the vent tube
21