CS110 Electric Field Meter
VoltDiff(Internal_RH,1,mV2500,5,True,0,250,0.1,0) 'Internal humidity
measure.
Changing of CS110 desiccant is recommended for internal relative humidity
values 80%.
FIGURE 14. Inside of CS110 case illustrating bracket
for holding desiccant
Replacement intervals less than once every six months for the 4
Unit (PN: 4905) desiccant pack within the sealed CS110 case
indicate a problem with the CS110 case seal or with the
desiccant packs being used.
NOTE
Procedure for Changing Desiccant:
1. Remove the CS110 case lid by unscrewing the captive screws that attach
the lid to the main body of the CS110.
38