CS110 Electric Field Meter
Rugged Construction: Ultra-reliable metallic ground connection to
reciprocating shutter (no wiping contact), brushless
stepper motor, powder-coated aluminum case,
Teflon insulators, and electro-polished 316L
stainless steel used for corrosion protection of
critical exposed metallic parts
Easy Maintenance: The stator is easily removed for cleaning (proper
cleaning does not invalidate calibration).
Instrument self-checking allows maintenance to be
performed on an as needed basis. The self-checking
also monitors internal humidity, insulator
cleanliness, and power supply voltage, and verifies
that CS110 components such as the charge
amplifier and shutter open/close are functioning
properly.
Operating
Temperature Range: -25° to 50°C standard, -40° to +85°C optional
RH Range: 0 to 100% RH
Dimensions: 15.2 x 15.2 x 43.2 cm (6 x 6 x 17 in)
Mounting: Vertical pipe 1.91 to 6.35 cm OD (0.75 to 2.5 in)
Weight: 4 kg (9 lb)