Appendix D. CR9000X Status Table
Field Name
Variable
Type
Description
CalAmpOffset(#) Integer
Array
Displays the Offset calibration factor for the different voltage ranges.
CalOffset(#) shows the calibration factor for the voltage ranges as depicted in
the following table. These values are updated at program compile time or
when a Calibrate or BiasComp instruction is encountered in the program, if the
program uses the measurement range.
# Value
Voltage Range
1 5000 mV
2 1000 mV
3 500 mV
4 500 mVX
5 200 mV
6 50 mV
CalBiasLo(#) Integer
Array
Displays the Offset calibration factor for the different voltage ranges.
CalOffset(#) shows the calibration factor for the voltages range as depicted in
the following table. These values are updated at program compile time or
when a Calibrate or BiasComp instruction is encountered in the program, if the
program uses the measurement range.
# Value
Voltage Range
1 5000 mV
2 1000 mV
3 500 mV
4 500 mVX
5 200 mV
6 50 mV
CalBiasHi(#) Integer
Array
Displays the Offset calibration factor for the different voltage ranges.
CalOffset(#) shows the calibration factor for the voltage ranges as depicted in
the following table. These values are updated at program compile time or
when a Calibrate or BiasComp instruction is encountered in the program, if the
program uses the measurement range.
# Value
Voltage Range
1 5000 mV
2 1000 mV
3 500 mV
4 500 mVX
5 200 mV
6 50 mV
D-4