Campbell-scientific CPEC200 Closed-Path Eddy-Covariance System Instrukcja Użytkownika Strona 95

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 112
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 94
Appendix F. Installing the AC/DC
Power Adapter Kit
The AC/DC Power Adapter Kit is configurable within the CPEC200 system
enclosure to allow the CPEC200 to be powered from AC mains power. A
peripheral mounting kit (pn 16987) is necessary to install the AC/DC adapter
into the CPEC200 system enclosure. The mounting kit includes a bracket, a
Velcro
®
strap, and the necessary nuts and screws.
The following steps describe the mounting procedure.
1. Install the peripheral mounting kit inside the CPEC200 system
enclosure by threading the Velcro
®
strap through the bracket.
2. Place the bracket at the backplate of the enclosure below the
datalogger and attach it with the screws included in the mounting kit
(see FIGURE F-1).
FIGURE F-1. Peripheral mounting kit installed in CPEC200 enclosure
3. Place the power supply on the mounting bracket and secure the power
supply’s cable to the back of the mounting bracket with a wire tie as
shown in FIGURE F-2.
F-1
Przeglądanie stron 94
1 2 ... 90 91 92 93 94 95 96 97 98 99 100 ... 111 112

Komentarze do niniejszej Instrukcji

Brak uwag