Campbell-scientific CPEC200 Closed-Path Eddy-Covariance System Instrukcja Użytkownika Strona 87

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 112
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 86
Appendix C. Output Variables
TABLE C-5. Values stored in table config_history
Variable Name
Units
Description
CO2_SPAN_PPM
µmol∙mol
-1
CO
2
mixing ratio in the CO
2
span cylinder
USE_DIFF_PRESS
Set = True to enable the EC155 differential pressure sensor
CHECK_ZERO
Set = True to enable automatic zero/span sequences
SET_ZERO
Set = True to set the zero during automatic zero/span sequences
CHECK_SPAN1
Set = True to check the CO
2
span during automatic zero/span
sequences
SET_SPAN_1
Set = True to set the CO
2
span during automatic zero/span
sequences
CHECK_SPAN2
Set = True to measure CO
2
span cylinder 2 during automatic
zero/span sequences
CHECK_SPAN3
Set = True to measure CO
2
span cylinder 3 during automatic
zero/span sequences
CHECK_SPAN4
Set = True to measure CO
2
span cylinder 4 during automatic
zero/span sequences
VALVE_MODULE
Compile switch: set = True to enable the use of the valve module
SCRUB_MODULE
Compile switch: set = True to enable the use of the scrub module
SaveAll_diagnostics
Compile switch: set = True to save additional diagnostic values in
the time series data table
Leaf_Wetness_Sensor
Compile switch: set = True to control intake heater voltage based
on a leaf wetness sensor
C-11
Przeglądanie stron 86
1 2 ... 82 83 84 85 86 87 88 89 90 91 92 ... 111 112

Komentarze do niniejszej Instrukcji

Brak uwag