CS547A Conductivity and Temperature Probe and A547 Interface
24: IF (X<=>F) (P89)
1: 3 X Loc [ Ct ]
2: 4 <
3: .474 F
4: 30 Then Do
25: Z=X*F (P37)
1: 3 X Loc [ Ct ]
2: .95031 F
3: 3 Z Loc [ Ct ]
26: Z=X+F (P34)
1: 3 X Loc [ Ct ]
2: -.00378 F
3: 3 Z Loc [ Ct ]
27: Else (P94)
28: Polynomial (P55)
1: 1 Reps
2: 3 X Loc [ Ct ]
3: 3 F(X) Loc [ Ct ]
4: -.02889 C0
5: .98614 C1
6: .02846 C2
7: .000000 C3
8: .000000 C4
9: .000000 C5
29: End (P95)
;This next program set will correct errors in the EC measurement resulting
;from temperature differences.
;
30: Temp (107) (P11)
1: 1 Reps
2: 3 SE Channel
3: 2 Excite all reps w/E2
4: 4 Loc [ Temp_degC ]
5: 1.0 Mult
6: 0.0 Offset
31: Z=X+F (P34)
1: 4 X Loc [ Temp_degC ]
2: -25 F
3: 6 Z Loc [ A ]
32: Z=X*F (P37)
1: 3 X Loc [ Ct ]
2: 100 F
3: 7 Z Loc [ Ct100 ]