CS511-L Dissolved Oxygen Probe
3. Pour clean water into the lower body and look for leakage around the
membrane. Dispose of the water, and if there is leakage, replace
membrane (see Section 8.1, Cleaning Probe and Replacing the
Membrane).
Water
4. Pour fresh electrolyte in the bottom cap and fill to the top of the cap.
5. Keep the probe upright with the cable pointed upwards (not sideways).
Screw the bottom cap onto the upper body until hand tight.
Excess electrolyte will leak out at the joint between the probe’s
cap and upper body.
NOTE