Campbell-scientific CR3000 Micrologger Instrukcja Użytkownika Strona 104

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 103
Section 7. Installation
104
Figure 41: DevConfig Deployment tab
Datalogger Sub-Tab
Serial Number displays the CR3000 serial number. This setting is set at the
factory and cannot be edited.
OS Version displays the operating system version that is in the CR3000.
Station Name displays the name that is set for this station. The default
station name is the CR3000 serial number.
PakBus® Address allows users to set the PakBus® address of the
datalogger. The allowable range is between 1 and 4094. Each PakBus®
device should have a unique PakBus® address. Addresses >3999 force other
PakBus® devices to respond regardless of their respective PakBus® settings.
See the PakBus® Networking Guide (available from Campbell Scientific) for
more information.
Security - See Security
(p. 70).
ComPorts Settings Sub-Tab
As shown in figure DevConfig Deployment | ComPorts Settings Tab
(p. 106), the
following settings are available for comports.
Read More! PakBus
®
Networking Guide available at www.campbellsci.com.
Selected Port specifies the datalogger serial port to which the beacon interval
and hello-setting values are applied.
Przeglądanie stron 103
1 2 ... 99 100 101 102 103 104 105 106 107 108 109 ... 589 590

Komentarze do niniejszej Instrukcji

Brak uwag