SECTION 3. INSTRUCTION SET BASICS
3-6
TABLE 3.9-1. Input/Output Instruction Memory and Execution Times
INSTRUCTION
INPUT
LOC.
PROG.
BYTES
EXECUTION TIME (ms)
MEASUREMENT RANGE
Code 0 1-4 or NA 5 Code 10 11-14 15 Code 20 21-24 25 Code 30 31-34 35
1 VOLT (SE) R 15 3.0 + 40.2 * R 4.6 + 5.2R 2.0 + 2.8R 1.6 + 20.6 * R 2.1 + 2.8R 1.8 + 2.6R 3.0 + 124.6 * R 12.8 + 13.5R 11.6 + 12.3R 2.3 + 142.4 * R 14.5 + 15.2R 12.1 + 12.8R
2 VOLT (DIFF) R 15 1.0 + 40.0 * R 0.6 + 9.1R 0.8 + 4.1R 2.2 + 20.5 * R 0.5 + 4.2R -0.6 + 4.1R 2.2 + 125.4 * R 0.6 + 25.8R 0.0 + 21.0R 3.5 + 142.4 * R 0.5 + 29.1R -0.2 + 24.3R
3 PULSE R 15 1.3 + 0.9R
4 EX-DEL-SE R 20 0.8 + 41.9 * R 4.7 + 5.3R 2.3 + 2.9R 1.4 + 22.5 * R 2.3 + 2.8R 2.1 + 2.7R 0.9 + 126.9 * R 13.1 + 13.6R 10.7 + 11.2R 2.1 + 143.4 * R 14.8 + 15.2R 12.3 + 12.9R
above w/o delay; w/delay: F + R(Tr + delay), where F is the fixed time and Tr is the time/rep from above.
(with delay) 0.8 + R *
(41.9+5*Delay)
1.4 + R *
(22.5+5*Delay)
0.9 + R *
(126.9+5*Delay)
2.1 + R *
(143.4+5*Delay)
5 AC HALF BR R 18 (-5.7) + 73.1 * R 6.9 + 9.6R 4.5 + 4.8R (-2.2) + 44.7 * R 4.5 + 4.7R 4.3 + 4.4R (-17.5) + 206.9 * R 15.2 + 26.3R 12.8 + 21.5R 4.1 + 229.4 * R 17.0 + 29.6R 14.5 + 24.8R
6 FULL BR R 18 (-3.9) + 98.1 * R 3.3 + 17.4R 3.3 + 7.8R (-4.3) + 54.3 * R 3.3 + 7.5R 3.3 + 6.7R (-16.3) + 268.9 * R 0.6 + 52.7R 0.7 + 43.0R 2.9 + 301.4 * R 1.9 + 59.2R 0.7 + 49.7R
7 3W HALF BR R 18 2.7 + 121.9 * R 6.6 + 21.3R 4.2 + 11.4R 19.4 + 64.5 * R 4.1 + 11.6R 3.9 + 10.6R 4.8 + 335.5 * R 13.6 + 56.4R 10.1 + 46.5R (-16.7) + 385.7 * R 15.3 + 63.1R 12.9 + 53.1R
8 EX-DEL-DIFF R 20 0.3 + 42.5 * R 4.7 + 5.3R 2.3 + 2.9R 4.5 + 22.7 * R 2.4 + 2.9R 2.1 + 2.7R 3.9 + 127.0 * R 13.1 + 13.6R 10.7 + 11.2R 3.8 + 143.8 * R 14.8 + 15.3R 12.3 + 12.9R
above w/o delay; w/delay: F + 2R(Tr + delay), where F is the fixed time and Tr is the time/rep from above.
(with delay) 0.3 +
2*R*(42.5+5*Delay)
4.5+2*R*(22.7+
5*Delay)
3.9+2*R*(127.0+5*
Delay)
3.8+2*R*(143.8+5
*Delay)
9 FULL BR-MEX R 19 (-6.0) + 327.0 * R 1.4 + 37.0R 1.5 + 17.5R (-7.0) + 85.0 * R 1.3 + 17.1R 1.6 + 15.3R (-8.7) + 475.0 * R -2.3 + 107.4R -2.2 + 87.9R 3.3 + 541.0 * R 3.4 + 118.9R -2.0 + 101.1R
V1 on range 5, V2 on: 1.6 + 27.2R 1.5 + 17.5R 1.6 + 17.3R 1.5 + 16.4R 0.3 + 62.4R -0.9 + 52.6R -0.3 + 69.1R 0.2 + 59.3R
V1 on range 15, V2 on: 1.5 + 26.2R 1.6 + 16.4R 1.6 + 16.2R 1.6 + 15.3R 0.3 + 61.3R -0.9 + 51.6R 0.3 + 68.1R 0.3 + 58.2R
10 BATT. VOLT 1 4 8.3
11 TEMP-(107) R 15 2.3 + 7.6R 14 + 48.4R 14.7 + 53.4R
12 RH (207) R 17 2.3 + 9.6R
13 TEMP-TC SE R 18 11.6 + 68.3 * R 8.5 + 9.2R 5.6 + 6.8R 11.9 + 49.2 * R 6.0 + 6.8R 5.2 + 6.7R (-2.8) + 138.5 * R 16.4 + 17.7R 13.4 + 15.2R 13.8 + 149.7 * R 18.0 + 19.4R 15.1 + 16.9R
14 TEMP-TC DIF R 18 11.5 + 69.3 * R 4.5 + 13.5R 4.2 + 8.3R 13.0 + 49.8 * R 4.3 + 8.6R 4.1 + 8.0R 215.8 + 35.0 * R 4.0 + 30.9R 4.6 + 25.1R 13.3 + 151.1 * R 4.2 + 34.2R 4.1 + 28.8R
15 SERIAL I/O R*#data 29 time in parameter 9
16 TEMP-RTD R 15 1.5 + 2.0R
17 TEMP-INTERNL 1 4 6.2
18 TIME 1 or 5 7 1.9 (options 0, 1, 2) 4.2 (option 3)
19 SIGNATURE 1 4 3600
20 PORT SET 0 6 11.3
21 PULSE PORT 0 5 1.2 + 10H; where H is hundredths of a second
22 EXCIT-DEL 0 9 0.5 + delay (ms)
23 BURST X 35
1.333R
# scans; where X = smpl before trigger + # chan
scans
24 CALIBRATION 19 4 2614, 25.0 when only saving results of automatic calibration.
25 READ PORTS 1 6 0.5
26 TIMER 1 or 0 4 0.4 to reset, 0.9 to load into location
27 PER.AVG. R 19
period of signal
(No. cycles + 1.5), or time limit + 2
28 VIB.WIRE R 21
170 + period of signal
(No. cycles + 1.5), or time limit + 172
29 INW PS9105 2 88 153
101 SDM-INT8 1-8 27
2.3 + 1.65 No.Values+averaging interval if used
102 SDM-SW8A R 16
4 + F
R; F=function time = 1 for state, 2 for counts or cycle, 0 for signature
103 SDM-AO4 0 6 1.1 + 0.8R
104 SDM-CD16 0 6 0.3 + 1.8R
105 SDI-12 REC X 18 sensor dependent; where X depends on SDI-12 sensor
106 SDI-12 SEN 0 10 ttt in parameter 2, this is sensor dependent
107 SDM-CSAT3 5R 7 2.5 + 2.4R
108 SDM-UDG01 1 15 155.4
109 SDMX50 0 4 1.7 + 31.1C; C = channel number
110 SDM TRIG 0 2 1.1
114 SET TIME 3-5 5 994.6
115 SET SDM BAUD 0 0
25 + 10
par 1
117 READ ID 0 4 0.8
Przeglądanie stron 59