CMP6-L, CMP11-L, and CMP21-L Pyranometers
7: Set Active Storage Area (P80)
1: 1 Final Storage Area 1
2: 101 Array ID
8: Real Time (P77)
1: 1220 Year,Day,Hour/Minute (midnight = 2400)
9: Average (P71)
1: 1 Reps
2: 3 Loc [ Solar_Wm2 ]
10: If time is (P92)
1: 0 Minutes (Seconds --) into a
2: 1440 Interval (same units as above)
3: 10 Set Output Flag High (Flag 0)
11: Set Active Storage Area (P80)
1: 1 Final Storage Area 1
2: 102 Array ID
12: Real Time (P77)
1: 1220 Year,Day,Hour/Minute (midnight = 2400)
13: Resolution (P78)
1: 1 High Resolution
14: Totalize (P72)
1: 1 Reps
2: 4 Loc [ Solar_MJ ]
15: Resolution (P78)
1: 0 Low Resolution
8. Maintenance
At regular intervals, physically inspect the pyranometer to ensure that:
Dome is free of dirt, condensation, and ice (see Section 8.1, Cleaning
Domes).
Desiccant granules are orange and opaque (see Section 8.2, Changing the
Desiccant).
Mounting is secure.
Pyranometer is level (if mounted horizontally).
Cables are in good condition.
22