Appendix J. AP200 Pump Replacement
FIGURE J-8. Pump side with inlet and outlet tubing connected
2. Reattach the pump to metal box with two self-tapping screws on the
back of the pump module electronics plate (FIGURE J-5).
Be careful to not pinch the fan wires under the pump and do not
overtighten the screws.
3. Reattach the pump connector to the pump electronics (FIGURE J-4).
4. Place the pump assembly back into the foam of the pump enclosure
(FIGURE J-3).
5. Hold the pump assembly securely while replacing the cover of the
pump enclosure. Make sure the fan does not slide back out of its hole
in the foam. Fasten the cover in place with the six screws as indicated
in FIGURE J-2.
Do not overtighten screws.
NOTE
NOTE
J-5