SI-111 Precision Infrared Radiometer
;{CR10X}
*Table 1 Program
01: 1 Execution Interval (seconds)
;Instruction string to measure the resistance of the thermistor and calculate the sensor body
;temperature. See the Instruction Manual for Campbell Scientific Model 109 Temperature Probe for
;details.
1: AC Half Bridge (P5)
1: 1 Reps
2: 25 2500 mV 60 Hz Rejection Range ;the range should at least match the excitation
3: 1 SE Channel
4: 1 Excite all reps w/Exchan 1
5: 2500 mV Excitation
6: 1 Loc [ mV_thrm ]
7: 1.0 Mult
8: 0.0 Offset
2: Z=1/X (P42)
1: 1 X Loc [ mV_thrm ]
2: 2 Z Loc [ 1_mV_thrm ]
3: Z=X+F (P34)
1: 2 X Loc [ 1_mV_thrm ]
2: -1.0 F
3: 3 Z Loc [ 2_mV_thrm ]
4: Z=X*F (P37)
1: 3 X Loc [ 2_mV_thrm ]
2: 24900 F
3: 4 Z Loc [ R_thrm ]
5: Z=LN(X) (P40)
1: 4 X Loc [ R_thrm ]
2: 5 Z Loc [ InR_thrm ]
6: Z=X*F (P37)
1: 5 X Loc [ InR_thrm ]
2: 0.001 F
3: 6 Z Loc [ Scaled_R ]
7: Polynomial (P55)
1: 1 Reps
2: 6 X Loc [ Scaled_R ]
3: 7 F(X) Loc [ SH_Coeffs ]
4: .001129 C0
5: .234108 C1
6: 0.0 C2
7: 87.7547 C3
8: 0.0 C4
9: 0.0 C5
8: Z=1/X (P42)
1: 7 X Loc [ SH_Coeffs ]
2: 8 Z Loc [ SB_Temp_K ]
6
Przeglądanie stron 11