Campbell-scientific CS526 ISFET pH Probe Instrukcja Użytkownika

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Urządzenia pomiarowe Campbell-scientific CS526 ISFET pH Probe. Campbell Scientific CS526 ISFET pH Probe User Manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj

Podsumowanie treści

Strona 1 - INSTRUCTION MANUAL

INSTRUCTION MANUAL CS526 ISFET pH Probe Revision: 4/14 Copyright © 2009- 2014 Campbell Scientific, Inc.

Strona 2

CS526 ISFET pH Probe 3. Initial Inspection Upon receipt of the CS526, inspect the packaging and contents for damage. File damage claims with the shi

Strona 3 - Limited Warranty

CS526 ISFET pH Probe 3. When Short Cut opens, select New Program. 4. Select Datalogger Model and Scan Interval (default of 5 seconds is OK for mos

Strona 4 - Assistance

CS526 ISFET pH Probe 5. Under the Available Sensors and Devices list, select the Sensors | Water | Quality folder. Select CS526 ISFET pH Probe. Cli

Strona 5 - Precautions

CS526 ISFET pH Probe 8. If LoggerNet, PC400, or PC200W is running on your PC, and the PC to datalogger connection is active, you can click Finish in

Strona 6

CS526 ISFET pH Probe Operating Temperature: 10° to 40°C Water Pressure: 0 to 700 kPa (0 to 101.5 psi) Power Requirement Source: 5 Vdc Load:

Strona 7 - Table of Contents

CS526 ISFET pH Probe 7.2 Datalogger Programming 7.2.1 Get Data Command The datalogger needs to send a “get data” serial command to the CS526 to get

Strona 8 - Figures

CS526 ISFET pH Probe 7.2.3 Instruction Descriptions 7.2.3.1 SerialOpen() Instruction The SerialOpen() instruction has the following syntax: SerialOp

Strona 9 - CS526 ISFET pH Probe

CS526 ISFET pH Probe 7.2.3.4 SplitStr() Instruction The SplitStr() instruction has the following syntax: SplitStr(SplitResult,SearchString,FilterStr

Strona 10 - 4. Quickstart

CS526 ISFET pH Probe Problem: Output signal is at its maximum value. Possible reasons: o Probe is not in fluid. o Chip is polluted. o Diaphragm is

Strona 11

CS526 ISFET pH Probe Clean the probe if any of the following occur: • Low slope • Drift • Instability of the reading • Slow calibration • Probe

Strona 13 - 6. Specifications

CS526 ISFET pH Probe 9.1.2.1 Cleaning Tips • To avoid scratches on the sensor surface, rinse the probe thoroughly using water before cleaning the pr

Strona 14 - 7. Operation

Appendix A. Importing Short Cut Code This tutorial shows: • How to import a Short Cut program into a program editor for additional refinement. • H

Strona 15 - 7.2 Datalogger Programming

Appendix A. Importing Short Cut Code A-2

Strona 16

Appendix B. Example Program The following is a CR1000 program that measures the CS526. This program assumes the CS526 is connected to COM1 (C1 / TX

Strona 17 - 8. Troubleshooting

Appendix B. Example Program B-2

Strona 18 - 9. Maintenance

Appendix C. Calibration This calibration process uses 7-pH and 4-pH buffer solutions. Protect the sensing chip from UV radiation during calibration

Strona 19 - 9.1.2 Cleaning Procedure

Appendix C. Calibration C.1 Example Calibration Program Following is a simple example program to facilitate the two-point calibration. Although thi

Strona 21

Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbe

Strona 22

Limited Warranty The CS526 ISFET pH Probe is warranted for six (6) months subject to this limited warranty: “Products manufactured by CSI are warrante

Strona 23 - Appendix B. Example Program

Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing

Strona 24

Precautions DANGER — MANY HAZARDS ARE ASSOCIATED WITH INSTALLING, USING, MAINTAINING, AND WORKING ON OR AROUND TRIPODS, TOWERS, AND ANY ATTACHMENTS TO

Strona 26

Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to spec

Strona 27

Table of Contents A.1 Importing Short Cut Code into a Program Editor ... A-1 B. Example Program ...

Strona 28

CS526 ISFET pH Probe 1. Introduction The CS526 ISFET pH Probe measures pH from 1 to 14 in aqueous or semi-solid solutions. It outputs TTL serial da